Send Message
Home All Cases

Imported lithium battery backup power supply scheme for anesthesia machine

I'm Online Chat Now

Imported lithium battery backup power supply scheme for anesthesia machine

October 21, 2020
Latest company case about Imported lithium battery backup power supply scheme for anesthesia machine

 

Anesthesia machine is to send anesthetics into the alveoli of patients through mechanical circuit, forming the partial pressure of anesthetic gas, which will directly inhibit the central nervous system and produce the effect of general anesthesia. The anesthesia machine belongs to a semi open anesthesia device. It is mainly composed of liquid medicine tank, ether regulating switch, folding bellows, suction and expiratory one-way valve and bellows.

 

The backup battery of anesthesia machine requires that it can work continuously for more than 8 hours under the condition of power failure, and the cycle life is required to reach 1000 times. Our company adopts imported Sanyo battery to meet the requirements of anesthesia machine.

 

Battery pack model GLD-14.8V-4.4AH
Model of single battery Sanyo lithium ion 18650 cell
Battery combination mode Three in series and two in parallel
Cell voltage 3.7V
Voltage after battery pack combination 14.8V (3.7V / section * 8 sections)
Battery charge and discharge voltage range after combination 12-16.8V
Cell capacity 2200mAh
Battery capacity after combination 4400mAh / section * 1 Section
Battery pack size 18*66*145mm
Maximum discharge current >2A
Instantaneous discharge current >4A
Maximum charging current >0.2-0.5C
Charge and discharge times > 1000 times

 

Lithium battery protection function of anesthesia machine

Lithium battery cell: imported Sanyo 18650 lithium ion battery cell is adopted;

Battery overcharge function: add a charger between P + and P - to charge the battery. When the voltage is charged to 16.8 ± 0.05V (it can be designed according to the requirements), the protection circuit will act and turn off the charging channel to realize overcharge protection;

Battery over discharge protection function: connect the load between P + and P - to discharge the battery. When the battery voltage drops to 12V ± 0.1V (it can be designed according to the requirements), the protection circuit will act and turn off the discharge path to realize the over discharge protection;

Short circuit protection function: when p+ and P- are short circuited, the protection circuit will act quickly within 5-50us to cut off the path and realize short-circuit protection;

Over current protection function: when the voltage between b-terminal and p-terminal reaches (0.15 ± 0.02V), the protection circuit will act quickly within 5-26ms to cut off the path and realize the over-current protection.

 

Temperature performance

Discharge temperature range: -10 ℃ - 60 ℃

Charging temperature range: 0 ℃ - 40 ℃

Storage temperature: 3 months: -20 ℃ - 45 ℃, 1 month: -20 ℃ - 60 ℃

Contact Details
Unitech Energy System Limited

Contact Person: Mrs. Olinda

Tel: +86-181 2856 1764

Fax: 86-755-23449332

Send your inquiry directly to us (0 / 3000)